AnalyteValueUnit95% C.I.NR/I/C
Al2.6mg/kg1.1 - 4.13I
Ba0.21mg/kg0.1 - 0.323I
Br0.38mg/kg0.31 - 0.458R
Ca149mg/kg139 - 15919R
Cd0.017mg/kg0.007 - 0.0276I
Cl570mg/kg460 - 68012R
Co0.0023mg/kg0.0014 - 0.00326I
Cs0.002mg/kg0.0004 - 0.00364I
Cu0.95mg/kg0.76 - 1.1420R
Fe4.1mg/kg3.4 - 4.815R
K1925mg/kg1790 - 206026R
Mg121mg/kg112 - 13012R
Mn2.06mg/kg1.94 - 2.1822R
Mo0.12mg/kg0.06 - 0.183I
Na2.6mg/kg1.7 - 3.510I
P592mg/kg506 - 6788R
Rb0.48mg/kg0.41 - 0.557R
Sb0.0026mg/kg0.0006 - 0.00464I
Zn2.53mg/kg2.2 - 2.8623R

(Value) Concentration calculated as a mean of the accepted laboratory means
(N) Number od accepted laboratory mean which are used to calculate the recommended or information values and their respective confidence intervals
(R/I/C) Classification assigned to the property value for analyte (Recommended/Information/Certified)

The values listed above were established on the basis of statistically valid results submitted by laboratories which had participated in one, or more, international intercomparison exercise organized by the Chemistry Unit, Agency’s Laboratories, Seibersdorf. The details concerning the criteria for qualification as a recommended or information value can be found in the respective reports which are available free of charge upon request. Orders for this material should be sent to AQCS, Seibersdorf.